General specifications | |
Power requirements | 100 V ac to 240 V ac, 47 Hz to 63 Hz |
Fuse | T2A 250 V ac |
Max power consumption | 100 W |
Operating ambient temperature range | 15 °C to 35 °C |
Storage temperature | -20 °C to 70 °C |
Relative humidity | Operating: <80 % to 30 °C, <70 % to 40 °C, <40 % to 50 °C |
Storage: <95 %, non-condensing. A power stabilization period of four days may be required after extended storage at high temperature and humidity. | |
Vibration | MIL-T-28800 |
Altitude (operation) | <2000 m |
Ingress protection | IEC 60529: IP20 |
Safety | IEC 61010-1, Installation Category II, Pollution degree 2 |
Weight (Chassis only) | 13 kg (28.5 lbs) |
Dimensions | Height: 147 mm (5.78 in) |
Width: 454 mm (17.79 in) | |
Depth 488 mm (19.2 in) | |
Rack mount dimensions | 3U-19 inch rack |
Warm up time | 15 minutes typical |
Control specifications | |
Control precision (dynamic mode) | PM200-BG2.5K range ± 0.005% range All other ranges ± 0.001% range |
Control turndown* | 10:1 (typical) |
Low control point | 1 kPa (0.15 psi) absolute |
Supply pressure requirements | |
Clean dry N2 or air – Industrial grade nitrogen, 99.5 %+ | |
Particulate contamination | ≤ 1.25 micrometer (50 microinches) |
Maximum moisture content | -50 °C dew point |
Maximum hydrocarbon content | 30 ppm |
Vacuum supply requirements (when operating near or sub-atmospheric) | |
>50 liters per minute capacity with Auto Vent feature | |
Appropriate protections for High Pressure Gauge work system exhaust gas will pass through the Vacuum supply system. | |
Interface/communications | |
Primary remote interfaces | IEEE, Ethernet, RS232, USB |
System connection | Supports interconnection of two or three systems |
Switch test connection | Standard 4 mm jack: Nominal 24 v dc isolated drive Maximum 30 V dc w.r.t. chassis ground |
Aux drivers | 4 external solenoid drivers 24 V dc drive (maximum drive 6 W continuous per channel) |
* Control turndown is defined as the relationship between the provided supply pressure and the appropriate supply pressure for the range. For example, a unit with a 7 MPa (1000 psi) and 700 kPa range (100 psi) with a supply pressure of 7.7 MPa (1100 psi) will provide control precision of 0.001% range because 7 MPa is 10 times greater than 700 kPa. A system with ranges of 20 MPa (3000 psi) and 700 kPa (100 psi) with supply pressure of 22 MPa (3300 psi) will have 0.001% range control precision on the 20 MPa range but only 0.003% control precision on the 700 kPa range. Control precision of 0.001% on the low range can be achieved by reducing the supply pressure. |
Model | Gauge mode range (SI units) | Absolute mode range (SI units) |
Gauge mode range (Imperial units) | Absolute mode range (Imperial units) | 1 year Specification | ||
Relative uncer- tainty (% Reading) |
Threshold uncer-tainty (% Span) |
Absolute mode adder (% Full scale) |
|||||
PM600-BG15K | -15 to 15 kPa | - | -60 to 60 inH₂0 | - | 0.01 % | 0.003 % | - |
PM600-G100K | 0 to 100 kPa | - | 0 to 15 psi | - | 0.01 % | 0.003 % | - |
PM600-G200K | 0 to 200 kPa | - | 0 to 30 psi | - | 0.01 % | 0.003 % | - |
PM600-A100K | -100 to 0 kPa | 6 to 100 kPa | -13.8 to 0 psi | 0.9 to 15 psi | 0.01 % | 0.003 % | 0.007 % |
PM600-A200K | -90 to 100 kPa | 10 to 200 kPa | -13.2 to 15 psi | 1.5 to 30 psi | 0.01 % | 0.003 % | 0.007 % |
PM600-A350K | -90 to 250 kPa | 10 to 350 kPa | -13.2 to 35 psi | 1.5 to 50 psi | 0.01 % | 0.003 % | 0.007 % |
PM600-A700K | -82 to 700 kPa | 18 to 700 kPa | -12.1 to 100 psi | 2.6 to 100 psi | 0.01 % | 0.003 % | 0.007 % |
PM600-A1.4M | -0.065 to 1.4 MPa | 0.035 to 1.4 MPa | -10 to 200 psi | 5 to 200 psi | 0.01 % | 0.003 % | 0.007 % |
PM600-A2M | -0.03 to 2 MPa | 0.07 to 2 MPa | -5 to 300 psi | 10 to 300 psi | 0.01 % | 0.003 % | 0.007 % |
PM600-A3.5M | -0.03 to 3.5 MPa | 0.07 to 3.5 MPa | -5 to 500 psi | 10 to 500 psi | 0.01 % | 0.003 % | 0.007 % |
PM600-A7M | 0 to 7 MPa | atmosphere to 7 MPa | 0 to 1000 psi | atmosphere to 1000 psi | 0.01 % | 0.003 % | 0.007 % |
PM600-A10M | 0 to 10 MPa | atmosphere to 10 MPa | 0 to 1500 psi | atmosphere to 1500 psi | 0.01 % | 0.003 % | 0.007 % |
PM600-A14M | 0 to 14 MPa | atmosphere to 14 MPa | 0 to 2000 psi | atmosphere to 2000 psi | 0.01 % | 0.003 % | 0.007 % |
PM600-A20M | 0 to 20 MPa | atmosphere to 20 MPa | 0 to 3000 psi | atmosphere to 3000 psi | 0.01 % | 0.003 % | 0.007 % |
BRM600-BA100K | - | 70 to 110 kPa | - | 10 to 16 psi | 0.01 % | - | - |
* 1-year specification is instrumental measurement uncertainty that includes linearity, hysteresis, resolution, reference uncertainty, 1-year stability, and temperature effects from 15 to 35 °C. The uncertainties are provided at 95% confidence, k=2, normally distributed. | |||||||
** Gauge mode uncertainty is the greater of the relative uncertainty and the threshold uncertainty. For absolute ranges used in gauge mode there is an additional uncertainty of ± 7 Pa for dynamic barometric compensation. When combined with other uncertainties this changes the threshold uncertainty for the PM600-A100K to 0.008% Span and for the PM600-A200K to 0.004% Span. | |||||||
** Absolute mode uncertainty is the greater of the relative uncertainty and the threshold uncertainty plus the absolute mode uncertainty adder for the lowest range PM600-AXXX module installed. For example, if a PM600-A200K and a PM600-A2M, the uncertainty at 2000 kPa absolute would be 0.2 kPa (0.01% * 2000 kPa) plus 0.014 kPa. |
Model |
Range (SI units) |
Range (Imperial units) |
Measurement mode |
Uncertainty (%FS) |
PM200-BG2.5K |
-2.5 to 2.5 kPa |
-10 to 10 inH₂0 |
gauge |
0.20 % |
PM200-BG35K |
-35 to 35 kPa |
-5 to 5 psi |
gauge |
0.05 % |
PM200-BG40K |
-40 to 40 kPa |
-6 to 6 psi |
gauge |
0.05 % |
PM200-A100K |
2 to 100 kPa |
0.3 to 15 psi |
absolute |
0.10 % |
PM200-BG100K |
-100 to 100 kPa |
-15 to 15 psi |
gauge |
0.02 % |
PM200-A200K |
2 to 200 kPa |
0.3 to 30 psi |
absolute |
0.10 % |
PM200-BG200K |
-100 to 200 kPa |
-15 to 30 psi |
gauge |
0.02 % |
PM200-BG250K |
-100 to 250 kPa |
-15 to 36 psi |
gauge |
0.02 % |
PM200-G400K |
0 to 400 kPa |
0 to 60 psi |
gauge |
0.02 % |
PM200-G700K |
0 to 700 kPa |
0 to 100 psi |
gauge |
0.02 % |
PM200-G1M |
0 to 1 MPa |
0 to 150 psi |
gauge |
0.02 % |
PM200-G1.4M |
0 to 1.4 MPa |
0 to 200 psi |
gauge |
0.02 % |
PM200-G2M |
0 to 2 MPa |
0 to 300 psi |
gauge |
0.02 % |
PM200-G2.5M |
0 to 2.5 MPa |
0 to 360 psi |
gauge |
0.02 % |
PM200-G3.5M |
0 to 3.5 MPa |
0 to 500 psi |
gauge |
0.02 % |
PM200-G4M |
0 to 4 MPa |
0 to 580 psi |
gauge |
0.02 % |
PM200-G7M |
0 to 7 MPa |
0 to 1000 psi |
gauge |
0.02 % |
PM200-G10M |
0 to 10 MPa |
0 to 1500 psi |
gauge |
0.02 % |
PM200-G14M |
0 to 14 MPa |
0 to 2000 psi |
gauge |
0.02 % |
PM200-G20M |
0 to 20 MPa |
0 to 3000 psi |
gauge |
0.02 % |
* Gauge mode modules (PM200-GXXX or PM200-BGXXX) with ranges of 100 kPa (15 psi) or greater will support absolute mode measurement when used with a Barometric Reference Module. |
Buy: Fluke 6270A - modular pressure controller, price: ---- euro
If you have any questions about this product feel free to contact with us by email: This email address is being protected from spambots. You need JavaScript enabled to view it. OR phone: +370 (5) 210 22 74.
By subscribing to our mailing list you will always be update with the latest news from us. We never spam.
JSC "Tek Know Baltic" Lithuania, Vilnius, Savanoriu ave 65A, LT03149
Tel: +370 (5) 210 22 74, Fax: +370 (5) 210 22 75, Email: info@tekknow.lt
© 2016, JSC "Tek Know Baltic". All rights reserved.