Savanoriu 65A, Vilnius, Lithuania +370 (5) 210 22 74 info@tekknow.lt
  • Differential pressure standard
  • Total uncertainty 40 ppm of reading +0.0013 psi (9 Pa)
  • Static line pressure range to 2900 psi (200 bar, 20 MPa)
  • Differential pressures to 30 psi (2100 mbar, 210kPa)
  • Hydraulic and pneumatic operation
  • Software provided for operation and data management
  • Extremely fast and easy to operate

The Model 2482 is a high precision standard that provides unsurpassed performance in the field of differential pressure metrology at high static line pressures. The Model 2482 continues the Fluke Calibration standard of excellence and innovation. Designed to calibrate virtually any pressure device at high static line pressures, the 2482 is a revolutionary breakthrough, providing a fast, easy and accurate solution to what has historically been a time consuming and technique dependent process.

At the heart of the Model 2482 is a patented large cross-sectional area triple-axis differential piston assembly with axial symmetry less than one millionth of a meter. This unique assembly eliminates the need for a second deadweight piston and provides identification of the differential pressure as it is directly proportional to the mass load.

In traditional twin-post and divider systems minute load adjustments are required to establish zero differential at each line pressure. This results in a process that is both technique dependent and time consuming. This elaborate zeroing process is eliminated in the Model 2482 through the use of a precision load cell, actually a digital force balance. The load cell supports the weight of the differential piston and serves as a null detector to ensure the line pressure balance point.

Establishing a differential pressure is then as simple as loading the weights determined by the supplied WinPrompt® software and activating the control function. Once the control system restores the balance reading to zero the desired differential pressure is applied to the test instrument.

The differential piston assembly is manufactured from tungsten carbide which provides high strength, durability and long-term stability. This material also provides a very low dependence on temperature, low distortion from pressure and undetectable hysteresis.

The Model 2482 is supplied with a precision mass set in a sturdy wooden storage case. The mass set contains various denominations that are trimmed to provide precise output pressures with standard increments to 0.5 inH2O (1.0 mbar). The Winprompt® software performs pressure-to-mass and mass-to-pressure calculations for the Model 2482 along with the corrections for environmental influences. The operator is also provided with a display of the pressure being generated by a mass load. The environmental conditions for humidity, barometric pressure and ambient temperature can be entered manually or the values can be acquired from a Model 2456-LEM Laboratory Environmental Monitor.

Included with the 2482 is the Controller software package that provides set up information, tared and line pressure displays, power settings for the temperature controllers, a graphic display of the tared readings as well as the control buttons for the major operations of the instrument.

The Model 2482 is a hydraulically operated piston gauge that allows for gas calibrations through two gas/oil interface chambers. A manual hand pump is provided for line pressure generation of either hydraulic or pneumatic pressures. For pneumatic operation a regulated gas supply is required.

General

Differential Pressure Range 0 to 30 psi (0 to 2100 mbar, 0 to 210 kPa)
Electrical Power Requirements 12V DC 5A max
120/240 VAC 50/60 Hz Universal power supply Included
Temperature Operating temperature 64-82°F (18-28°C)
Storage temperature 32-122°F (0-50°C)
Humidity Operating humidity 20-75% RH, non-condensing
Storage humidity 0-90% RH, non-condensing
Pressure Medium Operating medium for Piston/Cylinder, test pump and hydraulic test port: Dioctyl Sebacate (DOS) Operating medium for pneumatic test port: clean, dry, instrument grade gas (nitrogen recommended
Instrument Dimensions (h x w x d) 18 in x 15 in x 18 in (480 x 380 x 480 mm) Test Pump: 11 in x 11 in x 10 in (280 x 280 x 250 mm)
Instrument Weight 2482: 88 lb (40 kg)
Test Pump: 12 lb (5.5 kg)

Performance

Total Uncertainty
40 ppm of reading +0.0013 psi (9 Pa)

Total uncertainty is defined as the expanded uncertainty in pressure determined using the method described in the ISO “Guide to the Expression of Uncertainty in Measurement” and represents an approximate 95% level of confidence (2 sigma).

Static Pressure Accuracy 0.25% full scale
Long Term Stability Better than 10 ppm per year

Piston and Cylinder

Materials Tungsten carbide
Thermal Coefficient 9.2 x 10-6º C-1

Mass Set

Total mass 30 lbs (13.5 kg)
Max platter mass 9 lbs (4 kg)
Minimal pressure increment 0.018 psi (1.0 mbar, 125 Pa, 0.5 inH20)
0.5 inH2O ( 1.0 mbar)

Buy: Ruska 2482 - low pressure pump, price: ---- euro
If you have any questions about this product feel free to contact with us by email: This email address is being protected from spambots. You need JavaScript enabled to view it. OR phone: +370 (5) 210 22 74.

Do you want to get our newsletters?

By subscribing to our mailing list you will always be update with the latest news from us. We never spam.

JSC "Tek Know Baltic" Lithuania, Vilnius, Savanoriu ave 65A, LT03149

Tel: +370 (5) 210 22 74, Fax: +370 (5) 210 22 75, Email: info@tekknow.lt

© 2016, JSC "Tek Know Baltic". All rights reserved.